TM 11-6625-3087-12
CONTROL, INDICATOR,
FUNCTION
OR CONNECTOR
KEY
76
OFFSET push button
Used to enter, raise, lower, or display offset value stored in Level Meter. Press
to display present value, and if desired, enter anew value from -199.99dB to
+ 199.99dB using the numeric keys and -dB or +dB. New value can be stored
with offset on or off.
FULL SCALE push button
Used to enter, raise, lower, or display full scale value stored in Level Meter.
77
Press to display present value, and if desired, enter a new value using the
numeric keys and -dB or +dB. New value can be from -45dB to +25dB in 5dB
steps for 100dB range and from 120dB to +25dB in 5dB steps for 10dB range.
2-13