TM 11-6625-3087-12
Figure 2-2. Operator's Controls, Indicators, and Connectors, rear view.
CONTROL, INDICATOR,
FUNCTION
KEY
OR CONNECTOR
Line voltage switch
Used to select 100, 120, 220, or 240 volt operation. Two two-position
89
switches. Up-up for 100Vac operation, up-down for 120Vac operation, down-up
for 220Vac operation, and down-down for 240Vac operation.
Interface connector
Input and output connector for all external devices. Connector has 24 pins with
90
metric posts.
Interface control switch
91
Used to select tracking generator operating mode and address of AN/USM-490.
Seven-position DIP switch. Position one is interface troubleshooting test.
Position two sets tracking generator operating mode. Positions three thru seven
are for address settings.
Fo (0-32MHz) connector
92
Used for network analysis. Output frequency is 0 to 32.5 MHz and tracks the
Leave open for normal use. Connector mates with BNC plug.
1
93
10MHZ connector
Used for a signal reference to synchronize external devices to Level Meter
frequency reference. Output power is +8dBm at 50 Leave open for normal
operation. Connector mates with BNC plug.
2-16