TM 11-6625-3087-12
1. On Level Meter, press RECALL (7) then number 0 (6) push buttons.
The maximum input must not exceed 27dBm or damage to internal circuitry may
result
2. On Level Meter,
Select 75
input (8).
Connect rear panel Fo (0-32 .5MHz) output connector to front panel 75
BNC input using 75
cable supplied.
When using BNC input, use only 75
damage Level Meter.
3. On Level Meter, select desired UNIT of measurement (1).
4. On Level Meter,
Press 20Hz (4).
Press RDNG-->OFFSET (5).
5. Disconnect cable from front panel 75
BNC input.
6. On Level Meter, press OFFSET (3).
2-40