Verify the following condition exists on the Level Meter front panel:
d.
ON
MEASUREMENT/ENTRY display
ON
ON
CENTER indicator
KHz indicator
ON
AUTO CAL push button
ON
10dB push button
ON
AUTO push button
ON
ON
dBm push button
ON
CARRIER push button
HIGHER push button
ON
LO DIST push button
ON
10k shunted by 50pF push button
ON
ON
3100Hz push button
OFF
All other controls and indicators
e. If the above condition is obtained and no errors are displayed, the Level Meter is ready for use in one of the ten
NOTE
If the power cable has been disconnected for a period of time, the OVEN indicator will be
illuminated for approximately 20 minutes. Do NOT operate the Level Meter with the
OVEN indicator on. The measurements obtained will not be within specification.
2-7. STORE/RECALL PROCEDURES.
Up to nine front panel control settings and entry parameters can be stored for recall at a later date as follows:
1. To store a set-up:
Press the STORE (4) push button and the desired number from 1 to 9 using the numeric key board (6). The number
selected will flash in the MEASUREMENT/ENTRY display. If a set-up was previously stored at the location
selected (1-9), it will be erased and the new set-up will be stored.
2. To recall a set-up:
Press the RECALL (5) push button and the desired number from 1 to 9 using the numeric key board (6). The set-up
previously stored at that location will display.
3. To reset the Level Meter to the turn-on condition:
Press RECALL (5), and then O (6).
2-19