TM 11-6625-3087-12
MAINTENANCE PROCEDURES
Section IV.
DESCRIPTION
This procedure covers: Test.
INITIAL SETUP
NOTE
PRELIMINARY PROCEDURES:
TEST
1. Press ENTRY (1),100dB (2). Turn AUTO CAL (3) to off. Set full scale to 35dBm.
2. Press RECALL (4), . (decimal point) (5), CNTR--> FREQ (6), and number 9 (7) push buttons.
3. Verify that PASS is shown on the MEASUREMENT/ENTRY display (8)
4. Press MEAS CONT (9),
5. Press RECALL (4), . (decima point) (5), CNTR->FREQ (6), and number 8 (10) push buttons.
6. Verify that PASS is shown on the FREQUENCY/ENTRY display (11 ) for four tests, and musical tones are heard at the
speaker (12).
7. Press POWER SWITCH (13).
END OF TASK